The pressure transmitter series HDA 4800 has a very accurate and robust sensor cell with a thin-film strain gauge on a stainless steel membrane.

Outstanding technical data and robust construction make the HDA 4800 particularly suited to the field of test rig and diagnostic technology. It is also suitable for a broad range of applications in industry.

Since the accuracy of a pressure transmitter varies greatly with the temperature of the fluid, the instrument has excellent characteristics in thismore
Unit of Measure

Specifications

Model Code Description

HDA 4846-C-400-031

Type

Electronic Pressure Transmitter

Mechanical Connection Type

G1/4 A ISO 1179-2, Male

Input Burst Pressure

2000 bar

Mechanical Connection Size

G 1/4 A, Male

Output Signal

4 to 20 mA

Electrical Connection

Male M12 x 1, 4 Pole (Mating Connector Not Supplied)

BUn

EA

No of Conductor

3

Output Typical Accuracy According to DIN 16086, Terminal Based (Full Scale)

≤ ± 0.125 %

Output Typical Accuracy Best Fit Straight Line (B.F.S.L) (Full Scale)

≤ ± 0.06 %

Output Maximum Accuracy Best Fit Straight Line (B.F.S.L) (Full Scale)

≤ ± 0.125 %

Output Typical Temperature Compensation Zero Point (Full Scale)

≤ ± 0.005 %/°C

Output Maximum Temperature Compensation Zero Point (Full Scale)

≤ ± 0.01 %/°C

Output Typical Temperature Compensation Span (Full Scale)

≤ ± 0.005 %/°C

Output Maximum Temperature Compensation Span (Full Scale)

≤ ± 0.01 %/°C

Output Maximum Non-Linearity According to DIN 16086 Terminal Based (Full Scale)

≤ ± 0.15 %

Seal Material

Fluorocarbon Elastomer (FKM)

Material of Connection

Stainless Steel

Output Maximum Hysteresis (Full Scale)

≤ ± 0.1 %

Output Repeatability (Full Scale)

≤ ± 0.05 %

Output Rise Time

≤ ± 1 ms

Output Long-Term Drift Per Year (Full Scale)

≤ ± 0.1 %

CUL®US Mark

Certificate No.: E318391

Note for CUL®US Mark

Environmental conditions according to 1.4.2 UL 61010-1; C22.2 No. 61010-1

Direct Current (DC) Supply Voltage When Applied According to UL Specifications

12 to 30 V

Note for Direct Current (DC) Supply Voltage When Applied According to UL Specifications

Limited energy - According to 9.3 UL 61010; Class 2; UL 1310/1585; LPS UL 60950

Vibration Resistance According to DIN EN 60068-2-6 at 10 to 500 Hz

≤ 20 g

Shock Resistance According to DIN EN 60068-2-6 at 10 to 500 Hz Per 6 ms

≤ 100 g

Pin Connection 1

+UB

Pin Connection 2

No Connection

Pin Connection 3

0 Voltage

Pin Connection 4

Signal

Current Consumption

≤ 25 mA

Residual Ripple of Supply Voltage

≤ 5 %

Life Expectancy (0 to 100% Full Scale (FS))

> 10 Million Cycles

Input Overload Pressure

800 bar

Input Measuring Pressure Range

400 bar

Operating Temperature Range

-25 to 85 ºC

Note for Operating Temperature Range

-25 ºC (-13 ºF) with FKM seal, -40 ºC (-40 ºC) on request

Compensated Temperature Range

-25 to 85 ºC

Storage Temperature Range

-40 to 100 ºC

Medium Temperature Range

-25 to 100 ºC

Note for Medium Temperature Range

-25 ºC (-13 ºF) with FKM seal, -40 ºC (-40 ºC) on request

Protection Class According to DIN EN 605293

IP 67

Torque Value

20 N·m15 lb·ft

Industry Standards/Certifications

CE Mark EN 61000-6-1/2/3/4

Division

ELECTRONICS

Product Group Family

PRESSURE TRANSDUCERS

Product

PRESSURE TRANSDUCERS

Additional Information

N/A Since the accuracy of a pressure transmitter varies greatly with the temperature of the fluid, the instrument has excellent characteristics in this respect. By default, the output signals 4 to 20 mA, 0 to 10 V and 0 to 20 mA (source) are available.

Accessories

Accessories

Appropriate accessories, such as electrical female connectors, can be found in the Accessories brochure.